Naslov Stabilnost ortodontskih miniimplantata
Naslov (engleski) Stability of orthodontic mini-implants
Autor Domen Novšak
Mentor Stjepan Špalj (mentor)
Član povjerenstva Stjepan Špalj
Član povjerenstva Vladimir Ahel
Član povjerenstva Danko Bakarčić
Ustanova koja je dodijelila akademski / stručni stupanj Sveučilište u Rijeci Medicinski fakultet (Katedra za ortodonciju) Rijeka
Datum i država obrane 2012-11-30, Hrvatska
Znanstveno / umjetničko područje, polje i grana BIOMEDICINA I ZDRAVSTVO Dentalna medicina Ortodoncija
Sažetak Svrha: Cilj ovog istraživanja je bio utvrditi u kojoj mjeri biomehanički čimbenici - debljina kompakte, promjer i dužina miniimplantata, metoda uvrtanja te komponente vertikalne implantacijske sile utječu na primarnu stabilnost miniimplantata.
Materijali i metode: Testirano je šest tipova cilindričnih samobušećih miniimplantata Dual Top (Jeil Medical Corp., Seul, Koreja) promjera 1.4, 1.6 i 2.0 mm, dužine 6 i 8 mm uvrtanih ručno i strojno u uzorke svinjskih rebara bez pilot rupe. Mjereni su debljina kompakte pomoću pomične mjerke te okretni moment i vertikalna sila tijekom implantacije pomoću specijalnog dizajniranog uređaja povezanog s osciloskopom. Maksimalni implantacijski okretni moment rabljen je kao mjera primarne stabilnosti. U statističkoj analizi korištene su analiza kovarijance i Sidak post - hoc test te multipla hijerarhijska linearna regresijska analiza.
Rezultati: Statistički značajni prediktori bolje primarne stabilnosti ortodontskih miniimplantata su veća debljina kompakte (objašnjava 24.2% varijabilnosti), veći promjer miniimplantata (20.6%), ručni način uvrtanja (9.9%), veća dužina miniimplantata (3.7%), veća maksimalna vertikalna sila (2.2%) te manja vertikalna sila pri maksimalnom okretnom momentu (1.4%). Cijeli model opisuje 61.5% varijabilnosti primarne stabilnosti.
Zaključci: Na primarnu stabilnost miniimplantata najviše utječu debljina kompakte i promjer miniimplantata te način uvrtanja, dok dužina implantata i komponente vertikalne sile imaju manji utjecaj.
Sažetak (engleski) Objective: The aim of this study was to determine predictive values of biomechanical factors influencing primary stability of mini-implants: compacta thickness, mini-implant diameter and length, method of insertion, and components of vertical insertion force.
Materials and Methods: Six different types of cylindrical self-drilling mini-implants Dual Top (Jeil Medical Corp., Seoul, Korea) with diameters of 1.4, 1.6 and 2.0 mm and lengths of 6 and 8 mm were manually (hand-driven) and mechanically (machine-driven) inserted into the swine ribs bone samples without pre-drilling a guiding hole. Compacta thickness was measured with sliding caliper, while insertion torque and vertical insertion force were recorded with specially designed device connected with oscilloscope. Maximal insertion torque was used as a measure of primary stability. Analyses of covariance, Sidak post – hoc test and hierarchical multiple linear regression analysis were used in statistical analysis.
Results: Statistically significant predictors of better primary stability of orthodontic mini-implants are thicker compacta (explaining for 24.2% of variability), wider diameter of mini-implant (20.6%), manual insertion (9.9%), greater length of mini-implant (3.7%), higher maximal vertical insertion force (2.2%), and lower vertical force at maximal insertion torque (1.4%). A total of 61.5% of variability of primary stability is explained by this model.
Conclusions: Compacta thickness, mini-implant diameter, and manual insertion have higher impact on the primary stability of orthodontic mini-implants, while mini-implant length and components of vertical insertion force have lower impact.
Ključne riječi
miniimplantat
implantacijski okretni moment
promjer
dužina
vertikalna sila
način implantacije
Ključne riječi (engleski)
Mini-implant
Miniscrew
TAD
Insertion torque
Diameter
Length
Vertical force
Insertion mode
Jezik hrvatski
URN:NBN urn:nbn:hr:184:078403